Child DLA Diabetes CGM and Insulin Pump Evidence
When a child with type 1 diabetes uses a continuous glucose monitor, an insulin pump, or both, these technologies generate data that can support a DLA claim — but they also create a trap. Decision-makers sometimes assume the technology eliminates the need for parental care. Getting this right on the form is critical.
Why Technology Does Not Reduce Care Needs
A CGM (Dexcom G6/G7, FreeStyle Libre 2 Plus/3/3 Plus) reads interstitial glucose at regular intervals, which vary by model (for example, every 5 minutes for Dexcom G7 and every minute for Libre 3). It can raise configured high or low glucose alarms. A hybrid closed-loop system (CamAPS FX, Control-IQ, Medtronic 780G) adjusts insulin delivery based on CGM readings. Together, they automate some decisions.
What they cannot do:
- Treat a hypo. Some pump systems adjust or suspend insulin delivery in response to low CGM readings, but this does not treat a hypo. A child who can swallow safely may need fast-acting glucose and a recheck within 15 minutes; follow the child's hypo plan for any further treatment. An adult may need to respond, depending on the child's age, ability, and care plan.
- Fix a hardware failure. Sensors can drift, give compression lows (false readings from the child lying on the sensor), or lose signal. Pump cannulas can occlude, kink, or pull out. If a sensor fails at night, an adult may need to decide whether to replace it or switch to finger-prick testing.
- Count carbohydrates. A child may need help to weigh or estimate the carbs in a meal or snack and enter the number into the pump so it can calculate the bolus. The help required depends on the child's age and ability.
- Follow the treatment plan. Illness and exercise can affect glucose management. Follow the child's diabetes team's instructions on changes to targets, corrections, or basal rates; who makes those decisions depends on the child's age and ability.
- Change sites and sensors. Sensor wear time depends on the model (for example, up to 10 days for Dexcom G7 and up to 15 days for FreeStyle Libre 2 Plus or 3 Plus). Many standard pump infusion sets need changing every 2 to 3 days, while some extended sets are designed for up to 7 days; follow the product instructions. A young child may need help during a change.
How to Use CGM Data as Evidence
Download a 14-day AGP (Ambulatory Glucose Profile) report from LibreView, Clarity (Dexcom), or CareLink (Medtronic). This one-page summary shows:
- Time in range — what percentage of the day glucose is in range. A common target range is 3.9 to 10.0 mmol/L, but targets can be individualised. The percentage below range does not show whether a parent woke or intervened.
- Time below range — the percentage below 3.9 and below 3.0 mmol/L. The report shows glucose patterns, not whether an adult responded to an episode.
- Number of low events — some reports break this out per day. A low-event count does not establish how many episodes required treatment or parental attention; pair the report with a diary of what actually happened.
- Glucose variability — high variability shows unpredictable swings, but the metric alone does not establish that continual supervision is needed.
Attach the AGP to the claim pack. Highlight the overnight section and explain in the form what happens when each alarm fires.
Pump Data as Evidence
If your child uses an insulin pump, download the pump's summary report showing:
- Number of boluses per day — record which boluses required adult help, such as entering carbs or confirming a dose
- Number of alarms per day — occlusion alerts, low reservoir, sensor calibration requests
- Basal overrides — temporary rates or exercise modes, and who made the change
- Site change log — documenting changes at the interval required for that infusion set
A pump report showing 6 boluses a day and 4 alarms per night does not by itself establish frequent daytime attention or repeated night-time attention. Explain which events required an adult's attention, what they did, and how long it took.
Free Download
Get the Child DLA Evidence Checklist for Type 1 Diabetes
Everything in this article as a printable checklist — plus action plans and reference guides you can start using today.
Describing Technology on the Form
Do not write "my child wears a CGM and an insulin pump." Write what the technology requires you to do:
"My child wears a Dexcom G7 continuous glucose monitor that sends readings every 5 minutes. The attached report shows [X] overnight alarms over [period]. My diary records which alarms required an adult response and what I did, including checking the reading, giving fast-acting glucose when my child could swallow safely, and rechecking within 15 minutes. I followed my child's hypo plan for any further treatment. The CGM detects a low; it cannot treat it."
"My child's insulin pump delivers basal insulin automatically, but I must enter the carbohydrate count for every meal and snack so the pump can calculate the bolus dose. I also change the pump infusion set every 3 days, clean and prepare the new site, insert the cannula, prime the tubing, and check for bubbles. The pump set change takes about 15 minutes and my child finds it distressing."
The Decision-Maker's Mistake
If your claim is refused with the reasoning that "technology manages the condition," challenge it. The legal test is whether your child needs attention in connection with bodily functions, not whether a device has been prescribed. A device does not provide personal attention by another person; describe who actually provides the care.
Our Child DLA Type 1 Diabetes Guide includes a CGM and pump evidence summary template to help you present download data alongside your care diary, and explains how to counter the "technology reduces needs" argument at Mandatory Reconsideration or tribunal.